QUESTION

How do I connect my 3000SI to the LC control panel?


ANSWER

What follows is step by step instructions on how to install 3000SI sensors to an LC Control panel.


                                                                                      Channel 1 connection.

    •    Ensure power is disconnected from the system
  •    Connect green&yellow screen wire to LC Control panel earth stud
  •    Connect Red wire to Channel 1 pin 15
  •    Connect Blue wire to Channel 1 pin 14
  •    Configure Channel 1 jumper switches J1,J2,J4,J5,J8,J9 to on (closed)
  •    Configure Channel 1 jumper switches J3, J6, J7 to off (open)

   Subsequent sensors can be wired in a similar way:

   i.e. Channel 2 connection.

  •    Connect green&yellow screen wire to LC Control panel earth stud
  •    Connect Red wire to Channel 2 pin 27
  •    Connect Blue wire to Channel 2 pin 26
  •    Configure Channel 2 jumper switches J1,J2,J4,J5,J8,J9 to on (closed)
  •    Configure Channel 2 jumper switches J3, J6, J7 to off (open)
Step 3  Software Configuration

  • Connect power to the system and ensure detector head operates (text on LCD display, LED will flash once every six seconds approximately).
  • All information required for software configuration should be contained in the attached LC Software configuration and system calibration manual, a helpful 'map' of the menu structure can be found on page 4-9

Step 4. System Calibration

  • All information required for system calibration should be contained in the attached LC Software configuration and system calibration manual from page 4-16 onwards