We've recently had a few customers asking whether it would be possible for them to modify their older O2NE+ and 10yr O2NE to use the newly released 9100-2650/T oxygen sensors. This  is a relatively simple process and shouldn't be a problem for a suitably qualified technician to carry out. We have described the process in the attached document as well as demonstrated the modification of a device in the video below


A few tools will be necessary for carrying out the device modification, these are as follows:

  1. PPE: (Gloves, goggles)
  2. Drill
  3. 15mm spanner
  4. Knife
  5. Ruler
  6. Tweezers
  7. Deburring tool
  8. PZ1 screwdriver
  9. 16mm stepped bit
  10. 2mm bit


You will also need the following consumables:

  1. Replacement oxygen cell (9100-2650/T)
  2. Filter (1000-1600)


Once modification is complete, please view the Calibration solution for guidance on device calibration


If you require any guidance on any of these processes, please feel free to contact the Analox support team at: