1000 Series inc G21/22 (all variants)

1000 Series inc 1000, 5000, G21/22 (all variants)

1000: What depth can 1000 hyperbaric sensor work at?
QUESTION What depth can the hyperbaric sensor work at?   ANSWER Up to 60 Bar A
Thu, 13 Apr, 2017 at 1:20 PM
1000: How do I replace the dials on my Analox 1000?
QUESTION The dials on my Analox 1000 are in need of replacement, I have purchased new ones however I could do with some information on how to fit them ...
Thu, 13 Apr, 2017 at 1:20 PM
1000: What would happen if I apply too much gas pressure (over-pressure) to my 1000 sensor?
QUESTION What would happen if I apply too much pressure over my 1000?   ANSWER You will be given incorrectly high readings and eventuall...
Thu, 13 Apr, 2017 at 1:21 PM
1000: What diameter (size of tubing should I use to connect to the Analox 1000 flow adaptor?
QUESTION What diameter of tubing should I use to connect to the flow adaptor?   ANSWER We recommend that you use pipe with an outside diameter (O/...
Thu, 13 Apr, 2017 at 1:22 PM
G22: G22 doesn't reach 100% after calibration
QUESTION Our G22 analysers are calibrated using an 8% span gas and will 0 with cal gas, when we test the upper range with pure oxygen some of ...
Thu, 13 Apr, 2017 at 1:24 PM
5001: Negative reading at switch on?
QUESTION When I turn on my 5001 it shows a negative reading - is there a fault?   ANSWER No. The 5S2 sensor in the 5001 can take up to 40 seconds ...
Tue, 5 Jul, 2016 at 1:28 PM
5001: Will a presence of O2 in the span calibration gas have an effect?
QUESTION Does it matter if O2 is present in my span gas for my helium 5001?   ANSWER It shouldn't matter so long as no more than 20%...
Thu, 13 Apr, 2017 at 1:25 PM