The Ax60+ is primarily a breathing air device, as such, the CO2 sensor unit alarms are restricted to only trigger in the 'high-going' direction. If a user changes the alarm 'going' to 'low-going' a FLT16 will ensue. 


To clear this fault the following steps should be followed (be aware that only a short period of non-alarm time is available so you'll need to be quick!!):

  1. Power cycle the Ax60+ (turn it off and on again)
  2. Wait for the sensors display on the screen
  3. Press the Accept/Test button briefly to silence any alarms
  4. Press and hold the top two buttons (Cancel and cycle) until the top level menu is reached.
  5. Press the cycle button six times (until sensor config is shown)
  6. Press the Accept/Test button
  7. Press the cycle button twice (Alarm direction is shown)
  8. Press the Accept/Test button
  9. Select the sensor that had the alarm 'going' changed (press cycle to move the arrow selector over until the relevant sensor is identified and press the Accept/test button)
  10. Select the alarm that had the alarm 'going' changed (press cycle to move the arrow selector over until the relevant alarm is identified and press the Accept/test button) 
  11. Change the alarm direction to 'up' (press the cycle button until the arrow selector points at the up arrow and press the Accept/test button)
  12. Press the cancel button five times until the sensors display on screen
  13. Acknowledge any faults displayed
  14. Power cycle the unit