1] Remove the screws from the lid and remove the lid assembly from the base.


2] Remove the connector from the O2 sensor PCB.


3] Rotate the port connector anti clockwise and remove.


4] Remove the O2 cell and PCB assembly from the lid.


5] Remove the O2 cell from the PCB.


6] To fit a new O2 cell, repeat steps 1 to 6 in opposite sequence.


Please see page 36 in the service manual for the step by step guide with images.


Please watch the video below:

https://www.youtube.com/watch?v=8QM9uhqF6SU