Information regarding the use of the ACG+

ACG: COSHH regulations?
QUESTION Can the ACG be used to test the quality of breathing air used to protect personnel from hazards posed by a hazardous substance covered by the C...
Thu, 11 Oct, 2018 at 3:28 PM
ACG: Heliox / Trimix mixes?
QUESTION Can I use the ACG to measure Heliox or Trimix gas mixes? ANSWER No, the use of helium will produce inaccurate sensor readings. This ina...
Thu, 11 Oct, 2018 at 3:28 PM
ACG: Introductory video
This video gives a comprehensive introduction to the ACG.
Thu, 11 Oct, 2018 at 3:28 PM
ACG: Nitrox gas mixes?
QUESTION Can I use the ACG to measure Nitrox gas mixes? ANSWER Yes, the sensors are ranged to allow measurement of Nitrox mixes up to 50% O2. It shoul...
Mon, 10 Feb, 2020 at 9:31 AM
ACG: Recommended oil mist sample frequency?
QUESTION How often would you suggest I take an oil mist sample at the ACG’s sample support port? ANSWER The frequency of routine sample testi...
Thu, 11 Oct, 2018 at 3:28 PM
Can I use an ACG+ to monitor compressors using synthetic oil?
The ACG+ product page on the Analox Sensor Technology Website has all relevant product literature available for download. Take a look at the ACG+ oil mist ...
Thu, 11 Oct, 2018 at 3:08 PM
Is colorimetric tube sampling every three months not sufficient for testing breathing air?
In applications where three monthly tests are carried out, failed tests indicate that any compressed air produced since the last known good test is in que...
Thu, 14 Jan, 2021 at 1:29 PM
ACG+: Using colour metric tubes
QUESTION Can I use the sample support port for any type of colour metric tube ANSWER Yes, this feature adds flexibility to the user to use a var...
Tue, 5 Jul, 2016 at 2:19 PM
ACG: How do I install the ACG PC tool in Windows 10?
We've recently had a few customers reporting difficulties with using their Windows 10 systems to communicate with their ACG units. This solution aim...
Thu, 5 Jan, 2023 at 1:26 PM
ACG: How do I replace the PID particulate filter?
The ACG PID particulate filter should be replaced annually. This procedure is detailed in the ACG+ user manual.  We also have created an instruction...
Thu, 11 Oct, 2018 at 3:28 PM